The PRECICE nail system: The initial Kuwaiti experience
*Corresponding author: Mohammad M. Awad, Department of Pediatric Orthopedics and Deformity Unit, Kuwait City, Kuwait. al.mo.jawad@gmail.com
-
Received: ,
Accepted: ,
How to cite this article: Alsaifi SA, Hammady WK, Khaja AF, Awadh MM, El-Ashry SM, Awad M, et al. The PRECICE nail system: The initial Kuwaiti experience. J Musculoskelet Surg Res 2021;5:250-6.
Abstract
Objectives:
The PRECICE system is an implantable limb lengthening intramedullary nail with remotely magnetically controlled distractors indicated for limb length discrepancy (LLD) and short stature treatment. This study reports the initial experience of the Kuwaiti deformity correction unit in utilizing the PRECICE system.
Methods:
Ten patients (four females and six males) were included in this study. All cases were operated using the PRECICE nail system (five antegrade femoral nails, three retrograde femoral nails, and two tibial nails). All surgeries were performed during January 2019 to February 2020.
Results:
The mean age of participants was 20 years (12–33 years), with a 21.6 kg/m2 mean body mass index (17–28). LLD etiologies (mean LLD = 39 mm) were congenital (n = 2), developmental (n = 2), post-traumatic non-union (n = 1), post-traumatic malunion (n = 1), post-traumatic physeal arrest (n = 1), and post-deformity correction and lengthening of the contralateral side with circular frame (n = 3). The mean distraction rate was 0.97 mm/day (range: 0.75–1.2 mm/day). Mean lengthening was 39 mm (range: 20–60 mm). Healing was confirmed at 76 days on average (range: 50–120 days). All patients reached full consolidation to regenerate bone, normal alignment, and normal joint orientation. Antegrade femur lengthening was done in five patients. One patient with a previous knee fixed flexion deformity of 25° improved to a 5° lag of extension. No complications were observed during the lengthening procedures. All the patients were followed up for a minimum of 12 months.
Conclusion:
The PRECICE nail system was successful in lengthening cases with different etiologies, achieving target lengths without complications. All the patients had reported excellent functional outcomes.
Keywords
Intramedullary nail
Lengthening
Limb length discrepancy
PRECICE® nail
Deformity correction
Non-union
INTRODUCTION
Limb length discrepancy (LLD) can arise from various factors. These factors can be congenital or acquired.[1] Common acquired causes are seen in children and adolescents as growth plate arrests, mostly due to internal clinical entities or external direct forces.[2] Adults can acquire deformities due to osteomyelitis, traumas, malunions, or non-unions.[3-5] A. Codivilla initially described the modern-day concept of LLD and later Ilizarov further influenced the principles of limb equalization using external fixators.[6] His work was the cornerstone for the newer devices that evolved with time to face growing challenges in orthopedic surgery.[7,8]
The PRECICE system (NuVasive, San Diego, USA) is one of the newly developed technologies used in the treatment of LLD.[8] The nail itself was first introduced to patients in 2011 and gained the U.S Food and Drug Administration approval in the same year.[8] It started gaining world recognition soon after and has been further developed after addressing some pitfalls in the first version to produce a new improved version in 2013.[8] The nail is placed within the desired bone, and magnetically lengthened externally by the patient, nurse, or treating surgeon.[9,10] The rate and length of correction needed can be manipulated accordingly. It has been used successfully and reliably in many types of conditions like deformities, as well as LLD corrections and height lengthening.[10,11]
The PRECICE nail system is available in different lengths, diameters, strokes, angulations, and entry points.[12,13] These features accommodate for various needs and challenges that one might face in clinical practice. Although this versatility is useful to utilize, tailoring these various features can present quite a challenge in itself.
It is important to note that distraction osteogenesis through a fixator applied externally is an established technique for lower limb lengthening.[13-15] However, these treatments have had high complication rates, amounting to at most 3.2 complications per patient.[16,17] The pins or wires that penetrate the soft tissues trigger such complications as pain, pin-site infection, muscle transfixation, immobility, reduced joint movement, and scarring.[18,19] Removal of the external fixator puts the patient at risk of further complications, including malalignment or fracture.[20] Therefore, minimizing complications and improving patient comfort have necessitated the introduction of limb lengthening using fully implantable bone lengthening nails.[21] Magnetically-driven (PRECICE) or motorized FITBONE (Orthofix, Texas, USA) bone lengthening nails were made using mechanically-driven lengthening nails.[2,22,23] A couple of case-controlled studies have made comparisons between these nails and external fixation (13–15 patients), with the largest series involving 92 patients.[24-26] However, most reports of complications of PRECICE lengthening nails and FITBONE are small case series.[27-30] The popularity of motorized intramedullary lengthening nails has increased tremendously in recent years, leading to the hypothesis that the literature may now contain more standardized data on complications.
This is why the authors aimed at reporting their initial unit’s experience with the PRECICE nail. The aim was to investigate the functional outcomes and complications observed in the study group.
MATERIALS AND METHODS
The PRECICE nail system was used on ten patients seen in the deformity correction clinic in Al Razi Orthopaedic Hospital, Kuwait. This deformity correction unit is the only clinic in the country that offers this service. The data were prospectively collected during January 2019 to February 2020 as the patients underwent corrective surgeries using the PRECICE nailing system. The patients who had LLD that needed correction had various etiologies underlying their deformities.
The Arabic version of the Middle-East Lower Limb Score[31] was used for all the patients as a patient-reported outcome measure pre- and post-correction to validate the patient’s own subjective response.
Patients were followed up for 12 months post-surgical intervention to record complications and functional outcomes clinically and radiographically.
A total of ten patients (four females and six males) were included in this study. All the cases were corrected using the PRECICE nail system (five antegrade femoral nails, three retrograde femoral nails, and two tibial nails). Pre-operative planning for each case was performed regarding femoral or tibial lengthening, the nail entry point in the femoral cases (antegrade or retrograde), nail angulation (10° or straight), nail diameter (8.5, 10.7, and 12.5 mm), nail stroke (30, 50, or 80 mm), nail length, and osteotomy level. Patient factors (age, medical and surgical history, body mass index (BMI), LLD etiology, soft tissue size, and smoking status were accounted for in our planning.
RESULTS
Pre-operative target lengths were achieved. The mean age of the patients was 20 years (range, 12–33 years). The mean BMI of the patients was 21.6 kg/m2 (range, 17–28). The etiologies of LLD were congenital (Congenital short femur) (n = 2), developmental (a case of hypophosphatemia and a case of rickets) (n = 2), post-traumatic non-union (n = 1), post-traumatic malunion (n = 1), post-traumatic physeal arrest (n = 1), and post-deformity correction and lengthening of the contralateral side with circular frame (n = 3) [Table 1].
No | Age | Gender | BMI | LLD etiology | LLD in mm | Follow-up |
---|---|---|---|---|---|---|
Case 1 | 12 | F | 17 | Congenital short femur | 50 | 16 |
Case 2 | 16 | M | 20 | Post-traumatic physeal arrest | 40 | 16 |
Case 3 | 23 | M | 22 | Post-traumatic non-union | 50 | 12 |
Case 4 | 15 | M | 19 | Congenital short femur | 40 | 12 |
Case 5 | 20 | F | 28 | Developmental (Hypophosphatemia) | 60 | 12 |
Case 6 | 22 | M | 21 | Post-traumatic malunion | 30 | 12 |
Case 7 | 16 | M | 20 | Developmental (Rickets) | 20 | 12 |
Case 8 | 33 | F | 23 | Post-deformity correction and lengthening of contralateral femur with circular frame | 30 | 12 |
Case 9 | 19 | F | 22 | Post-deformity correction and lengthening of contralateral tibia with circular frame | 30 | 12 |
Case10 | 24 | M | 24 | Post-deformity correction and lengthening of contralateral tibia with circular frame | 40 | 12 |
Mean | 20 | 21.6 | 39 | 12.8 |
The mean latency period (time between the osteotomy and starting the distraction osteogenesis) was 7.8 days (range, 5–10 days). The mean distraction rate was 0.97 mm/day (range, 0.75–1.2 mm/day). Mean lengthening was 39 mm (range, 20–60 mm). Healing was confirmed at a mean of 76 days (range, 50–120) days. The mean time to full weight-bearing was 14.7 weeks (range, 10–23 weeks). All patients reached full consolidation, normal alignment, and normal joint orientation by the end of the study [Table 2]. The knee range of motion was maintained in five patients (antegrade femur lengthening). One patient with a previous knee fixed flexion deformity of 25° improved to 5° of extension lag (retrograde femur lengthening). Two patients lost 20° of flexion during the early post-operative period and 5° of extension (retrograde femur lengthening). This loss was regained later with extensive physiotherapy. One patient needed a pre-distraction technique (defined by the difference of leg length determined by radiographic images)[7] for the nail before the procedure to consider enough nail length for passing the non-union site of the previous trauma. No complications regarding pain during lengthening, failure of distraction, or hardware failure were reported [Figures 1-6].
No | Latency period | Distraction rate | Healing in (days) | FWB (days) | MELLS Pre-operative | MELLS Post-operative |
---|---|---|---|---|---|---|
Case 1 | 5 | 1.2 | 110 | 20 | 39 (Good) | 10 (Excellent) |
Case 2 | 10 | 0.75 | 70 | 14 | 63 (Fair) | 10 (Excellent) |
Case 3 | 10 | 0.75 | 80 | 15 | 45 (Good | 6 (Excellent) |
Case 4 | 7 | 1 | 70 | 14 | 58 (Good) | 12 (Excellent) |
Case 5 | 10 | 1 | 120 | 23 | 89 (Poor) | 17(Excellent) |
Case 6 | 7 | 1 | 60 | 13 | 70 (Fair) | 12 (Excellent) |
Case 7 | 5 | 1 | 50 | 10 | 76 (Poor) | 10 (Excellent) |
Case 8 | 7 | 1 | 60 | 12 | 76 (Poor) | 4 (Excellent) |
Case 9 | 7 | 1 | 60 | 12 | 33 (Good) | 4 (Excellent) |
Case 10 | 10 | 1 | 80 | 14 | 41 (Good) | 4 (Excellent) |
Mean | 7.8 | 0.97 | 76 | 14.7 | 59 | 9 |
DISCUSSION
Although the PRECICE nail system is relatively expensive, this high price has dissuaded surgeons from using it in their clinical practice.[10] Many centers have published their experiences on a small number of patients using the PRECICE nail system.[10,11] The results in the literature provided so far are promising, yet long-term complications of this system remain unknown.[12] However, when compared to the complications of other intramedullary systems, as well as the complications reported from circular frames, the PRECICE system has been found to have less device-related complications.[32-34] This has been reflected well by our patient’s satisfaction with the surgery and their compliance with the post-operative requirements.[34] Although the literature has reported nail distraction failures due to the embedded magnet,[12] this study did not note such a problem. Distraction-related severe pain was not reported in this study as well. This was in keeping with the literature.[35-38] The range of motion of the cohort in this study improved well without difficulties due to the hospital’s aggressive physiotherapy regimen. There was no delay in bone consolidation in all of the cohort of this study. This finding, too, was consistent with other similar studies in the literature.[12,34,35,39]
One of the major advantages of the PRECICE system is the lengthening mechanism, which is controlled externally, as well as the control of the distraction rate.[40,41] Schiedel et al. in their study found a discrepancy between the length measured radiologically and the length displayed on the external remote controller (ERC) monitor. This occurred in ten cases out of 23 in their study, while the regenerate bone was 10% shorter on average than anticipated. They have suggested possible reasons for that, as there might be an inadvertent displacement of the ERC during the lengthening procedure or the lengthening performance in an outpatient setting by the patient in the absence of the clinician. It is also worth knowing that the PRECICE system should not be applied in patients whose BMI exceeds 35. This is because the soft tissue thickness causes an increase in the distance between the ERC and the nail. This increasing distance may interfere with the rotations of the nail through the magnet.[41]
The previous versions of internal lengthening devices have had diverse mechanisms of operations, such as a linear actuator with an external electronic controller,[42] a spring and ratchet system powered by voluntary limb movement (rotations),[43,44] as well as a roller-clutch threaded rod assembly powered by rotations of the limb in intramedullary skeletal kinetic distractor (ISKD, Orthofix Inc, Lewisville, TX, USA).[45]
The mechanism notwithstanding, the primary cause of complications appears to be inaccurate control of distraction for clinical use.[46] The manufacturer claims that nail distraction can be controlled accurately through a customized programmable ERC.[47] It is, however, worth knowing that a magnet-powered internal lengthening appliance is a novel technology, with limited literature as per its clinical efficacy.[46-49]
The present study had a few limitations. First, the study sample size was small; however, this is attributed to the fact that this nail system has just been made available in the country. Second, the short follow-up period is another limitation of this study; however, it is shared by many other studies done on the PRECICE system because it is a relatively recent development. Another major limitation in this study is reflected in the different etiologies and the large age spectrum. The authors cannot establish full validity of the results due to these pitfalls, but the results still account as a reference and valuable addition to the literature. The cost of the PRECICE nail is evidently much higher than the use of external fixators and circular frame models. In this study, we did not address the cost-effectiveness of the PRECICE nail compared to other treatment modalities. Another noteworthy consideration is the possibility of pseudotumor formation in titanium implants. Surgeons using this devise should be vigilant of such a concern.
CONCLUSION
The PRECICE nail system is a valid and useful tool for LLD correction surgeries. The patients who participated in this study had an excellent outcome. Therefore, we recommend that this would be the treatment of choice for the indicated cases.
AUTHORS’ CONTRIBUTIONS
SS conceived the idea for this project. All authors contributed equally to this research. All authors made substantial contributions to the conception and design, data acquisition, analysis and interpretation. AK was involved in drafting the manuscript and making critical revisions for important intellectual content. All authors have critically reviewed and approved the final draft and are responsible for the manuscript’s content and similarity index.
ETHICAL APPROVAL
Ethical approval was obtained from the Ministry of Health, Kuwait, Research and publication office. Committee Reference Number: 2019/1065 (May 14, 2019).
Declaration of patient consent
The authors certify that they have obtained all appropriate patients consent forms. In the form, the patients have given their consent for their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.
Financial support and sponsorship
This study did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Conflicts of interest
There are no conflicts of interest.
References
- Practical aspects of posttrauma reconstruction with an intramedullary lengthening nail. J Orthop Trauma. 2017;31(Suppl 2):S20-4.
- [CrossRef] [PubMed] [Google Scholar]
- Precision of the PRECICE internal bone lengthening nail. Clin Orthop Relat Res. 2014;472:3869-78.
- [CrossRef] [PubMed] [Google Scholar]
- PRECICE intramedullary limb lengthening system. Expert Rev Med Devices. 2015;12:231-49.
- [CrossRef] [PubMed] [Google Scholar]
- How precise is the PRECICE compared to the ISKD in intramedullary limb lengthening? Acta Orthop. 2014;85:293-8.
- [CrossRef] [PubMed] [Google Scholar]
- A brief history of limb lengthening. J Pediatr Orthop. 2017;37:S1-8.
- [CrossRef] [PubMed] [Google Scholar]
- Outcomes following femoral lengthening. Bone Joint J. 2016;98B:1382-8.
- [CrossRef] [PubMed] [Google Scholar]
- Treatment of post-traumatic femoral discrepancy with PRECICE magnetic-powered intramedullary lengthening nails. J Orthop Trauma. 2017;31:369-74.
- [CrossRef] [PubMed] [Google Scholar]
- Humerus lengthening with the PRECICE internal lengthening nail. J Pediatr Orthop. 2017;37:e296-300.
- [CrossRef] [PubMed] [Google Scholar]
- Stature lengthening using the PRECICE intramedullary lengthening nail. Tech Orthop. 2015;30:167-82.
- [CrossRef] [Google Scholar]
- Two consecutive limb lengthenings with the same PRECICE nail: A technical note. Strategies Trauma Limb Reconstr. 2018;13:199-204.
- [CrossRef] [PubMed] [Google Scholar]
- Humeral Lengthening with the PRECICE magnetic lengthening nail. HSS J. 2017;13:217-23.
- [CrossRef] [PubMed] [Google Scholar]
- PRECICE® magnetically-driven, telescopic, intramedullary lengthening nail: Pre-clinical testing and first 30 patients. SICOT J. 2017;3:19.
- [CrossRef] [PubMed] [Google Scholar]
- Evaluation of the first experience of intramedullary nail lengthening using PRECICE® in a South African limb lengthening and reconstruction unit. S Afr Orthop J. 2016;15:67-71.
- [CrossRef] [Google Scholar]
- Limb lengthening by callus distraction (callotasis) J Pediatr Orthop. 1987;7:129-34.
- [CrossRef] [PubMed] [Google Scholar]
- Current techniques of limb lengthening. J Pediatr Orthop. 1988;8:73-92.
- [CrossRef] [PubMed] [Google Scholar]
- Clinical application of the tension-stress effect for limb lengthening. Clin Orthop Relat Res. 1990;250:8-26.
- [CrossRef] [Google Scholar]
- Limb lengthening by callus distraction: Complications in 53 cases operated 1980-1991. Acta Orthop. 1994;65:447-55.
- [CrossRef] [PubMed] [Google Scholar]
- Distraction osteogenesis of the lower extremity with use of monolateral external fixation: A study of two hundred and sixty-one femora and tibiae. J Bone Joint Surg Am. 1998;80:793-806.
- [CrossRef] [PubMed] [Google Scholar]
- Complications of Albizzia femoral lengthening nail: An analysis of 36 cases. J Pediatr Orthop B. 2012;21:394-9.
- [CrossRef] [PubMed] [Google Scholar]
- Patient satisfaction after limb lengthening with internal and external devices. J Surg Orthop Adv. 2015;24:174-9.
- [CrossRef] [PubMed] [Google Scholar]
- Fracture after distraction osteogenesis. J Bone Joint Surg Br. 2000;82:659-65.
- [CrossRef] [Google Scholar]
- The intramedullary skeletal kinetic distractor (ISKD): First clinical results of a new intramedullary nail for lengthening of the femur and tibia. Injury. 2001;32(Suppl 4):SD129-39.
- [CrossRef] [Google Scholar]
- Limb lengthening by implantable limb lengthening devices. Tech Orthop. 2014;2:72-85.
- [CrossRef] [Google Scholar]
- Internal lengthening device for congenital femoral deficiency and fibular hemimelia. Clin Orthop Relat Res. 2014;472:3860-8.
- [CrossRef] [PubMed] [Google Scholar]
- Lengthening in congenital femoral deficiency: A comparison of circular external fixation and a motorized intramedullary nail. J Bone Joint Surg Am. 2015;97:1432-40.
- [CrossRef] [PubMed] [Google Scholar]
- Femoral lengthening with a motorized intramedullary nail: A matched-pair comparison with external ring fixator lengthening in 30 cases. Acta Orthop. 2015;86:248-56.
- [CrossRef] [PubMed] [Google Scholar]
- Femoral lengthening using the PRECICE intramedullary limb-lengthening system. Bone Joint J. 2019;101B:1168-76.
- [CrossRef] [PubMed] [Google Scholar]
- Leg lengthening with a motorized nail in adolescents: An alternative to external fixators? Clin Orthop Relat Res. 2008;466:189-97.
- [CrossRef] [PubMed] [Google Scholar]
- Functional results of lower extremity lengthening by motorized intramedullary nails. Acta Orthoped Traumatol Turc. 2012;46:42-9.
- [CrossRef] [PubMed] [Google Scholar]
- Evaluation of the first experience of intramedullary nail lengthening using PRECICE in a South African limb lengthening and reconstruction unit. S Afr Orthop J. 2016;15:67-7.
- [CrossRef] [Google Scholar]
- Translation, adaptation and validation of the middle east lower limb score (MELLS) into Arabic. Int J Innov Res Med Sci. 2021;5:488-92.
- [CrossRef] [Google Scholar]
- Femoral lengthening over an intramedullary nail. A matched-case comparison with Ilizarov femoral lengthening. J Bone Joint Surg Am. 1997;79:1464-80.
- [CrossRef] [PubMed] [Google Scholar]
- Intramedullary skeletal kinetic distractor in the treatment of leg length discrepancy-a review of 16 cases and analysis of complications. J Orthop Trauma. 2012;26:e138-44.
- [CrossRef] [PubMed] [Google Scholar]
- Femoral lengthening and deformity correction using the Fitbone motorized lengthening nail. J Orthop Sci. 2015;20:149-54.
- [CrossRef] [PubMed] [Google Scholar]
- Use of a magnetic bone nail for lengthening of the femur and tibia. J Orthop Surg. 2016;24:374-8.
- [CrossRef] [PubMed] [Google Scholar]
- A retrieval analysis of the PRECICE intramedullary limb lengthening system. Bone Joint Res. 2018;7:476-84.
- [CrossRef] [PubMed] [Google Scholar]
- The PRECICE magnetic IM compression nail for long bone nonunions: A preliminary report. Arch Orthop Trauma Surg. 2019;139:1551-60.
- [CrossRef] [PubMed] [Google Scholar]
- Retrograde magnetic internal lengthening nail for acute femoral deformity correction and limb lengthening. Expert Rev Med Devices. 2017;14:811-20.
- [CrossRef] [PubMed] [Google Scholar]
- Surgical techniques for lengthening and deformity correction of the femur with lengthening nails. Tech Orthop. 2015;30:183-8.
- [CrossRef] [Google Scholar]
- How precise is the PRECISE lengthening nail? The west of Scotland experience. Orthop Proc. 2017;99:6.
- [Google Scholar]
- Motorized intramedullary nail for management of limb-length discrepancy and deformity. J Am Acad Orthop Surg. 2014;22:403-9.
- [CrossRef] [PubMed] [Google Scholar]
- How precise is the PRECICE compared to the ISKD in intramedullary limb lengthening? Reliability and safety in 26 procedures. Acta Orthop. 2014;85:293-8.
- [CrossRef] [PubMed] [Google Scholar]
- A fully implantable motorized intramedullary nail for limb lengthening and bone transport. Clin Orthop Relat Res. 1997;343:135-43.
- [CrossRef] [Google Scholar]
- Implantable devices for lengthening the femur without external drive mechanisms. Med Tekh. 1984;2:44-9.
- [CrossRef] [Google Scholar]
- Mechanical characterization of a totally intramedullary gradual elongation nail. Clin Orthop Relat Res. 1997;337:281-90.
- [CrossRef] [PubMed] [Google Scholar]
- The intramedullary skeletal kinetic distractor (ISKD): First clinical results of a new intramedullary nail for lengthening of the femur and tibia. Injury. 2001;32:SD129-39.
- [CrossRef] [Google Scholar]
- PRECICE Nail System: Ellipse Technologies. 2020. United States: Ellipse Technologies Inc; Available from: http://www.ellipse-tech.com/?q=ip [Last accessed on 2020 May 07]
- [Google Scholar]
- Magnetically controlled growing rods for severe spinal curvature in young children: A prospective case series. Lancet. 2012;379:1967-74.
- [CrossRef] [Google Scholar]
- A novel device for long bone osteodistraction: Description of device and case series. Plast Reconstr Surg. 2012;130:418e-22e.
- [CrossRef] [PubMed] [Google Scholar]