Effect of manual vancomycin loading on the bending strength and flexural modulus of bone cement
Article History
-
Received: ,
Accepted: ,
,
Abstract
Objectives:
Manual addition of antibiotics when prefabricated antibiotic-loaded bone cement is not available is possible but should be performed with caution due to the effects of the antibiotic on the mechanical properties of polymethyl methacrylate (PMMA). This study investigates differences in bending strength and modulus and morphology of bone cement resulting from two variations in manual vancomycin loading.
Methods:
Three groups of bone cement – control without antibiotic, group with coarsely mixed 1 g vancomycin, and group with finely mixed 1 g vancomycin/40 g bone cement were tested for bending strength and bending modulus according to ISO5833:2002. The broken surfaces of the tested samples were analyzed using scanning electron microscopy.
Results:
Both methods of manual vancomycin loading caused a significant decrease in bending strength but not in bending modulus. The bending strength of the group with finely mixed vancomycin was significantly above the ISO5833:2002 threshold of 50 MPa. The bending strength of the group with coarsely mixed vancomycin did not differ significantly from the minimum set by the standard, with 4 of 9 specimens failing to meet the minimum requirement per ISO standards. The bending modulus exceeded 1800 MPa, with no significant differences among groups. The addition of vancomycin increased microporosity and the incidence of microcracks in the cement samples.
Conclusion:
Manual loading of finely mixed vancomycin causes an acceptable decrease in the tested mechanical properties of bone cement and can be considered for use in arthroplasty procedures. Additional testing of compressive strength and fatigue could provide a sufficient mechanical profile of these formulations and a better understanding of whether manually loaded properties of PMMA with vancomycin are mechanically safe for implant fixation.
Keywords
How to cite this article: Popovska D, Avramov N, Boev I, Velkovski V, Bogojevska-Doksevska M, Todorova T. Effect of manual vancomycin loading on the bending strength and flexural modulus of bone cement. J Musculoskelet Surg Res. 2026;10:623-8. doi: 10.25259/JMSR_192_2026
INTRODUCTION
The practice of using antibiotic-loaded bone cement (ALBC) for primary or secondary infection prevention during joint arthroplasty procedures is well established, though still debated. Therefore, it is used according to different guidelines in different parts of the world.[1-3] Manual addition of antibiotics when prefabricated ALBC is unavailable is possible but should be performed with caution due to the antibiotics’ effects on the mechanical properties of polymethyl methacrylate (PMMA). This effect depends on the type and dose of the added antibiotic.[4]
The oldest method of bone cement preparation is manual mixing of the powder and liquid components. This method allows for the entrapment of air bubbles, which, along with monomer evaporation, causes porosity of the cement mixture. In vitro, porosity has been shown to decrease mechanical properties and to contribute to fatigue failure of bone cement.[5-7] Vacuum mixing has been introduced as a method to reduce cement porosity and improve its mechanical performance.[5,8] However, these in vitro results do not translate to clinical results. For example, low-viscosity cements, which have lower porosity, yield inferior clinical results for femoral stem fixation compared with high-viscosity cements.[9] The results from the Swedish hip registry also showed that implantation of the femoral component with low-viscosity cement mixed with vacuum was associated with a higher risk for revision in the first 4 years.[10] In ALBC, porosity is a factor that positively influences antibiotic elution.[11]
Various studies report differing results regarding mechanical performance, antibiotic elution, and antimicrobial activity when different mixing methods for manual antibiotic addition were employed.[5,12,13] This study investigated the differences in bending strength, bending modulus, and morphology of bone cement with two variations of manual vancomycin loading. We used commercially available bone cement and an antibiotic for intravenous use, as well as instruments used for routine bone cement preparation during surgery at the Clinic for Orthopedic Diseases in (Skopje, Republic of North Macedonia), with the intention to replicate the usual conditions of cement preparation with manual antibiotic loading during surgical procedures.
MATERIALS AND METHODS
We performed a 4-point bending test to determine bending strength and bending modulus in accordance with the international standard ISO5833:2002[14] and conducted scanning electron microscopy (SEM) analysis of bone cement with manually added vancomycin.
SmartSet™ ENDURANCE™ MV Medium Viscosity Bone Cement (DePuy Synthes/Warsaw, IN) was used to prepare cement samples. The cement from a 40 g standard package was loaded with 1g vancomycin in the form of commercially available lyophilized vancomycin hydrochloride intended for intravenous administration (Vankopol 500 mg, Polifarma, Istanbul, Türkiye). This vancomycin is in powder form and is composed of large, flaky granules. Three groups of samples were produced: PMMA control samples containing only bone cement, which were mixed according to the manufacturer’s instructions, and groups V1 and V2, each containing 1g of vancomycin. Cement mixing, sample preparation, and sample storage until testing were performed at 22°C and 50% relative humidity.
Group V1 with coarsely mixed vancomycin was prepared by pouring the powder component of the bone cement into a bowl, adding the vancomycin, mixing both with a spatula for 30 s to break down large antibiotic granules, and then adding the liquid component.
Group V2, with finely mixed vancomycin, was produced by first pouring the vancomycin, then grinding it with a pestle until it turned into a fine powder. Then the powder component of the bone cement was added, and both were manually mixed with a spatula for 3 min. After this process, the liquid component was added.
After adding the liquid component, the bone cement mixture in all groups was manually mixed with a spatula until it reached a doughy consistency, leaving no trace on a sterile, non-powdered, dry latex glove.
Cement samples with length 75 (±0.1) mm, width 10 (±0.1) mm, and thickness 3.3 (±0.1) mm were produced according to the procedure described in ISO5833:2002 Implants for surgery – Acrylic resin cements international standard Annex F: Determination of bending modulus and bending strength of polymerized cement.[14] Molds were made from polytetrafluoroethylene and filled with cement dough, clamped, and the cement was left to set for 1h before removing the cement samples [Figure 1a and b]. After removal, the samples were fine-ground to the above-specified dimensions.

Eleven or twelve samples were produced from each group; samples with large irregularities were discarded, and finally, 9 samples from each group were tested using a 4-point bending test 24 h after production. The test was performed using a Shimadzu AGS-X 10 kN materials testing machine (Shimadzu Corporation, Kyoto, Japan) with a speed range of 0.001–1000 mm/min [Figure 1c and d]. The test was performed at a crosshead rate of 5 mm/min. Deflection was measured with a deflection sensor of 10 nm precision. Deflection for 15N and 50N loads was recorded for the calculation of the bending modulus, and the force required for sample breaking was recorded for the calculation of the bending strength.
Morphological analysis of the broken surface of two samples of each group after the 4-point bending test was performed with a scanning electron microscope TESCAN VEGA3 LMU with a back-scattered electron detector in high vacuum. The sample surface was sputter-coated with gold.
The results were examined with one-sample t-tests, Levene’s test for equality of variances, and analysis of variance (ANOVA). We used an alpha level of 0.05 for all statistical tests.
RESULTS
4-point bending tests
The average values for bending modulus and bending strength are shown in Table 1. Bending modulus was higher than the minimum of 1800 MPa required in the standard for all examined groups, with an increase of 0.13% for V1 and 1.7% for V2 compared to PMMA. There was homogeneity of variances between groups, as assessed by Levene’s test for equality of variances (F = 0.744, p = 0.49). There were no statistically significant differences between group means as determined by one-way ANOVA (F [2,24] = 0.266, p = 0.769).
| Bending modulus (MPa) | Bending strength (MPa) | ||||||
|---|---|---|---|---|---|---|---|
| Group | Average | Range | SD | Average | Range | SD | Difference from minimum (50 MPa) p-value* |
| PMMA | 2556.1 | 2291.4–2730.0 | 128.6 | 65.2 | 59.4–70.8 | 3.8 | <0.001 |
| V 1 | 2559.4 | 2428.9–2693.7 | 93.1 | 51.7 | 33.6–64.5 | 11.2 | 0.340 |
| V 2 | 2599.4 | 2312.2–2937.7 | 184.3 | 59.2 | 51.8–69.1 | 4.9 | <0.001 |
The average bending strength across all groups exceeded the standard’s minimum of 50 MPa. PMMA and V2 had average bending strengths significantly above 50 MPa, whereas V1 showed no significant difference from this value. All 18 PMMA and V2 samples tested had bending strength above 50 MPa. Four of 9 samples from the V1 group had bending strength below the 50 MPa threshold.
Group V1 showed a 20.7% reduction in bending strength relative to PMMA, whereas V2 showed a 9.2% reduction. Levene’s test indicated unequal variances (F = 4.027, p = 0.031), with group V1 having significantly higher variability than PMMA (p = 0.045) as opposed to variability between PMMA and group V2 (p = 0.98). There was a statistically significant difference in average bending strength between the three groups as determined by one-way ANOVA (F [2,24] = 6.643, p = 0 005). Post hoc analysis using the Bonferroni method indicated that the bending strength of PMMA was significantly higher than that of V1 (p = 0.009) and V2 (p = 0.015). The difference between V1 and V2 did not reach statistical significance (p = 0.11).
Scanning electron microscopy analysis
SEM analysis showed a somewhat more irregular surface on samples containing vancomycin than on PMMA. Figure 2 depicts the surface of samples of all groups at magnifications of ×30 and ×200. The white structures scattered across the sample surfaces are agglomerates of the radiopacifier barium sulfate, as confirmed by energy-dispersive X-ray spectroscopy. There is an area of pronounced porosity on the surface of the V1 sample, as shown in the enlarged image V1 200×. The other sample examined from this group showed similar irregularities.

Higher magnifications revealed the structure of the polymer and the antibiotic particles, with multiple micropores and microcracks on the broken surfaces of the samples. Microcracks were detected spreading from micropores, connecting neighboring voids, or attaching to antibiotic particles [Figure 3]. Samples containing vancomycin exhibited more pronounced microporosity than PMMA samples, with isolated zones of microporosity within the polymer structure of V1 and V2 samples [Figure 4]. This occurrence was not widespread, and such dense microporosity was not detected in the PMMA samples.


DISCUSSION
Studies show that manually loading bone cement with antibiotics can yield samples with good antibiotic elution and mechanical properties comparable to those of commercial preparations. Frew et al. compared commercially available bone cement with gentamycin and vancomycin (Copal G+V) with gentamycin-loaded cement (Copal R+G), which they manually loaded with 2 g of vancomycin.[15] They found that manual vancomycin loading significantly increased elution of both antibiotics. Another study that examined gentamycin and tobramycin elution from bone cement prepared with four different methods of mixing (antibiotic suspension in the liquid phase; pouring of antibiotic and powder phase and immediate mixing with the liquid phase; manual mixing of antibiotic and powder phase and then pouring of liquid phase; mixer mixing of antibiotic and powder phase) found neither significant advantage of any of these mixing methods nor any definitive difference between these manually prepared formulations and commercially available ALBC with the same antibiotic composition.[16] In a study comparing manual, mechanical and industrial mixing of gentamycin and cement powder before manual mixing with the liquid component, Lewis et al. found no significant difference in bending strength, modulus, fracture toughness, fatigue, antibiotic elution, mean particle size, and the morphology of the blended cement powders between the 3 mixing methods, although they conclude that mechanical mixing with a mixer in the operating room may produce more consistent mixture compared to manual mixing.[17]
In this study, we compared certain mechanical and morphological characteristics of bone cement with those of bone cement containing manually added commercially available vancomycin. The addition of vancomycin at a dose of 1g/40g PMMA caused a significant decrease in cement bending strength, regardless of the mixing method. These results confirmed the results of previous studies examining the effect of vancomycin on the mechanical properties of bone cement.[18,19] In these studies, however, samples with 1g vancomycin/40g PMMA had bending strength that was not significantly different from 50 MPa. In our study, samples prepared by grinding and meticulous mixing of vancomycin had significantly higher bending strength than the minimum of 50 MPa specified by the international standard. This group exhibited low variability, nearly identical to that of PMMA, and all tested samples had bending strength values well above 50 MPa. The samples with coarsely mixed vancomycin had bending strength values that were not significantly higher than 50 MPa. The rough mixing of vancomycin produced samples with more unpredictable behavior, greater variability, and bending strength below the required minimum. There were no significant differences between groups for the bending modulus.
Laine et al. examined three different vacuum mixing methods of 1 g vancomycin in 40g PMMA (mixing presieved vancomycin in the cement powder, presieved vancomycin in the liquid component, and nonsieved vancomycin in the powder).[13] They hypothesized that sieving would result in a more homogeneous distribution of antibiotics and improved mechanical properties of the cement. They found that although samples with nonsieved vancomycin appeared to have greater porosity, their mechanical properties did not differ significantly from those of the control group, whereas the presieved groups performed worse in mechanical testing. They explain their findings with the theory that sieving caused homogeneous weakening of the polymer, whereas nonsieved samples had heterogeneous zones of weakened and intact polymer, which performed better in mechanical testing. In our study, we found a seemingly greater porosity in samples with coarsely mixed vancomycin. These samples performed worse in mechanical testing. We also observed greater porosity across all vancomycin samples than in PMMA. There were small micropores in the polymer structure that were independent of the antibiotic-filled voids, implying that vancomycin might interfere with PMMA polymerization to some extent. This microporosity may be eliminated by vacuum mixing.
Kim et al. found macropores with dimensions above 1 mm when adding vancomycin at doses above 1g in PMMA and mixing with vacuum.[18] In another study by the same group, they examined the effect of telavancin loading and found that higher antibiotic concentrations led to greater porosity and decreased fracture toughness. In the latter study, they found that pore-to-pore interactions influence fracture propagation.[20] We also found microcracks on the fracture surface that connected neighboring antibiotic particles or pores.
For preparing the cement mixture, we used materials and instruments readily available in any operating room where joint replacement is performed. We strictly followed the instructions of ISO5833:2002 for sample manufacturing and testing procedures. The standard requires testing of 5 samples. We tested 9 samples from each group, a larger sample size than in most studies examining the mechanical properties of PMMA. However, in this study, we were unable to test compressive strength, which is specified in the same standard; fatigue per ISO16402: Implants for surgery – Acrylic resin cement – Flexural fatigue testing of acrylic resin cements used in orthopedics; or other mechanical properties that would provide a more complete insight into the mechanical behavior of the tested ALBC formulations. Testing of compressive strength and fatigue warrants further investigation and could provide a sufficient mechanical profile of these formulations and a better understanding of whether manually loaded PMMA with vancomycin is mechanically safe for implant fixation. We did not have the option to measure the porosity of the samples, so our estimate of the higher porosity of ALBC samples is subjective.
CONCLUSION
According to this study’s findings, loading Smart Set Endurance™ MV with 1g of vancomycin reduces its bending strength. If this is done by grinding the granules to a fine powder and mixing the powder component for a sufficiently long period (3 min), this decrease is acceptable compared to coarse mixing. Therefore, the extra few minutes spent preparing the cement may be very important for the longevity of the implant.
Authors' contributions:
DP: Conceived and designed the study, conducted research, and wrote the initial and final draft. NA: Performed mechanical testing, software modifications, and data interpretation and analysis. IB: Provided resources, conducted research, and performed data analysis for SEM investigations. VV: Provided resources and analyzed and interpreted data. MBD performed statistical analysis and participated in initial manuscript drafting. TT: Collected data and performed a literature review. All authors have critically reviewed and approved the final draft and are responsible for the manuscript’s content and similarity index.
Ethical approval:
Institutional Review Board approval is not required as the study does not include research performed on humans or animals, biologic material or patient data of any kind.
Declaration of patient consent:
Patient’s consent is not required as there are no patients in this study.
Use of artificial intelligence (AI)-assisted technology for manuscript preparation:
The authors confirm that there was no use of AI-assisted technology for assisting in the writing or editing of the manuscript and no images were manipulated using AI.
Conflicts of interest:
There are no conflicting relationships or activities.
Financial support: This study did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
References
- Antibiotic-loaded bone cement is associated with a reduction of the risk of revision of total knee arthroplasty: Analysis of the catalan arthroplasty register. Knee Surg Sports Traumatol Arthrosc. 2025;33:354-63.
- [CrossRef] [PubMed] [Google Scholar]
- The use of antibiotic-loaded bone cement and systemic antibiotic prophylactic use in 2,971,357 primary total knee arthroplasties from 2010 to 2020: An international register-based observational study among countries in Africa, Europe, North America, and Oceania. Acta Orthop. 2023;94:416-25.
- [CrossRef] [PubMed] [Google Scholar]
- Evidence base and practice variation in acute care processes for knee and hip arthroplasty surgeries. PLoS One. 2017;12:e0180090.
- [CrossRef] [PubMed] [Google Scholar]
- Evaluation of the effect of ciprofloxacin and vancomycin on mechanical properties of PMMA cement; a preliminary study on molecular weight. Sci Rep. 2020;10:3981.
- [CrossRef] [PubMed] [Google Scholar]
- Effect of vacuum mixing on the mechanical properties of antibiotic-impregnated polymethylmethacrylate bone cement. J Biomed Mater Res. 1990;24:573-80.
- [CrossRef] [PubMed] [Google Scholar]
- The influence of antibiotics on the fatigue life of acrylic bone cement. J Bone Joint Surg Am. 2006;88(Suppl 4):148-55.
- [CrossRef] [PubMed] [Google Scholar]
- A fractographic investigation of PMMA bone cement focusing on the relationship between porosity reduction and increased fatigue life. J Biomed Mater Res. 1992;26:651-62.
- [CrossRef] [PubMed] [Google Scholar]
- Bone cement porosity in vacuum mixing systems In: Walenkamp GH, Murry DW, eds. Bone Cements and Cementing Technique. Berlin, Heidelberg: Springer; 2001. p. :81-95.
- [CrossRef] [Google Scholar]
- The effect of the type of cement on early revision of Charnley total hip prostheses. A review of eight thousand five hundred and seventy-nine primary arthroplasties from the Norwegian Arthroplasty Register. J Bone Joint Surg Am. 1995;77:1543-50.
- [CrossRef] [PubMed] [Google Scholar]
- How outcome studies have changed total hip arthroplasty practices in Sweden. Clin Orthop Relat Res. 1997;344:44-60.
- [CrossRef] [Google Scholar]
- Influence of antibiotic-loaded acrylic bone cement composition on drug release behavior and mechanism. Polymers (Basel). 2021;13:2240.
- [CrossRef] [PubMed] [Google Scholar]
- Mixing method affects elution and strength of high-dose ALBC: A pilot study. Clin Orthop Relat Res. 2012;470:2677-83.
- [CrossRef] [PubMed] [Google Scholar]
- Effects of mixing techniques on vancomycin-impregnated polymethylmethacrylate. J Arthroplasty. 2011;26:1562-6.
- [CrossRef] [PubMed] [Google Scholar]
- International Organization for Standardization ISO 5833:2002(en) Implants for Surgery-Acrylic Resin Cements. 2002 Available from: https://www.iso.org/standard/30980.html [Last accessed on 2021 Sep 27]
- [Google Scholar]
- Comparison of the elution properties of commercially available gentamicin and bone cement containing vancomycin with 'home-made' preparations. Bone Joint J. 2017;99-B:73-7.
- [CrossRef] [PubMed] [Google Scholar]
- Hand-mixed and premixed antibiotic-loaded bone cement have similar homogeneity. Clin Orthop Relat Res. 2009;467:1693-8.
- [CrossRef] [PubMed] [Google Scholar]
- Influence of the method of blending an antibiotic powder with an acrylic bone cement powder on physical, mechanical, and thermal properties of the cured cement. Biomaterials. 2005;26:4317-25.
- [CrossRef] [PubMed] [Google Scholar]
- Mechanical, elution, and antibacterial properties of simplex bone cement loaded with vancomycin. J Mech Behav Biomed Mater. 2020;103:103588.
- [CrossRef] [PubMed] [Google Scholar]
- Mechanical effects of the use of vancomycin and meropenem in acrylic bone cement. Acta Orthop. 2006;77:617-21.
- [CrossRef] [PubMed] [Google Scholar]
- Influence of porosity on fracture toughness and fracture behavior of antibiotic-loaded PMMA bone cement. J Biomech Eng. 2022;144:11006.
- [CrossRef] [PubMed] [Google Scholar]